dets/AUS3_322_00

Leakage

Timestamp Leakage (µA)
Unknow 1.000

Dead layer

Current json parameters is 663nm front dead layer and 4um on the back.

Run 1862

Re-investigation of thickness using the IFA009 setup. The relevant files are triple-alpha-SU_1862.

Done using GeometryCalibrator daf0d42 and DCalib 61956b

Position fitting

Starting from sorted files with a 100keV tolerance.

Hitpattern -l 1000 triple-alpha-SU_1862_*
IsotropicFitter -d setup/setup.json -i hist/h_*1862* -n SU -p chits/hit_SU -V

This gives

Center: (-0.475001, -0.588802, -38.0176)
 Err: (0.0250138, 0.0242563, 0.0717518)

Fitting dead layer

HistBuilder setup/setup.json "unpacked_data/triple-alpha-*1862*" SU hist/dl_1862
DCalib -k source_simple.json -d SU -r 12 -f -i hist/dl_1862

Results:

3182.69 keV peak gives: 570.603 ± 18.5829 nm
5105.5 keV peak gives: 595.419 ± 36.6334 nm
5144.3 keV peak gives: 594.737 ± 36.8391 nm
5156.59 keV peak gives: 594.512 ± 36.9043 nm
5762.64 keV peak gives: 596.141 ± 34.4475 nm
5804.96 keV peak gives: 594.723 ± 34.6365 nm
Total: 584.665 ± 12.1372
Fits